Project title of the project activity or programme of activities (PoA) for which issuance is requested | Quimobásicos HFC Recovery and Decomposition Project (0151) |
---|---|
DOE that requested for issuance and date of request | TÜV SÜD (07 Jul 11) |
Please indicate, in accordance with paragraphs 65 of the CDM modalities and procedures, for which reason(s) you request review | |
Incompetence | |
Please indicate reasons for the request for review | |
All equipment (except the flow meters) should be recalibrated monthly as per the applicable methodology(page 10): “All of the measurement instruments are to be recalibrated monthly per internationally accepted procedures except for the HFC 23 flow meters whose recalibration frequency is weekly to reduce the error level”. However, the revised monitoring plan (page 19) states that the calibration of the instrument used to measure the purity of the HFC23 waste gas should be calibrated every year and verified every 2 months. Paragraph 200 of the VVM (version 1.2) states that “the DOE shall verify that the validated monitoring plan is in accordance with the approved methodology applied by the proposed CDM project activity.” The DOE is requested to clarify why it did not raise a FAR in the verification report requiring the PP to revise the monitoring plan in order to align it to the applicable methodology before the next issuance request. | |
Date | 31 Aug 11 |