Info Report Check
Submission incomplete:
The calibration of measuring equipment has to be conducted at the frequency specified in the registered monitoring plan.


The Monitoring Report and Verification Report indicate that an error factor was applied because of a delayed calibration. However:

- It is not clear in the Monitoring Report neither in the Verification Report whether the value of error factor applied (2%) is the value obtained after calibration, since the accuracy of steam flow meter is 1.5%, as per the registered PDD.

- In addition, during March 2020 there was a delay in the calibration of the steam flow meter and no conservative approach has been applied during this period.

- There is a calculation error of the parameter Qsteam,y for March 2020 in the ER spreadsheet (Tab: 'Generation Details; Cell C98).